2.1
Monitoring
Requirement
2.1.1 In
accordance with the approved EM&A Manuals, 24-hour Total Suspended
Particulates (TSP) monitoring at the designated air quality monitoring station is
required. Impact 24-hour TSP monitoring should be carried out at least once
every 6 days. In case of complaints, 1-hour TSP monitoring should be carried
out at least 3 times per 6 days when the highest dust impacts are likely to
occur.
2.2
Monitoring
Methodology
2.2.1 24-hour TSP air quality monitoring
HVS
Installation
The
following guidelines were adopted during the installation of HVS:
n Sufficient
support is provided to secure the samplers against gusty wind.
n No two
samplers are placed less than 2 meters apart.
n The distance
between the sampler and an obstacle, such as buildings, is at least twice the
height that the obstacle protrudes above the sampler.
n A minimum of
2 meters of separation from walls, parapets and penthouses is required for
rooftop samples.
n A minimum of
2 meters separation from any supporting structure, measured horizontally is
required.
n No furnaces
or incineration flues are nearby.
n Airflow
around the samplers is unrestricted.
n The samplers
are more than 20 meters from the drip line.
n Any wire fence
and gate, to protect the sampler, should not cause any obstruction during
monitoring.
Filters
Preparation
Fiberglass
filters (provided by the HOKLAS accredited laboratory) shall be used (Note:
these filters have a collection efficiency of larger than 99% for particles of
0.3 µm diameter). A HOKLAS accredited laboratory (ALS Technichem (HK) Pty Ltd.)
is responsible for the preparation of 24-hr conditioned and pre-weighed filter
papers for monitoring team.
All
filters are equilibrated in the conditioning environment for 24 hours before
weighing. The conditioning environment temperature is around 25°C and not
variable by more than ±3°C; the relative humidity (RH) is < 50% and not
variable by more than ±5%. A convenient working RH is 40%.
Operating / Analytical
Procedures
Operating
/ analytical procedures for the air quality monitoring are highlighted as
follows:
n Prior to the
commencement of the dust sampling, the flow rate of the HVS are properly set
(between 0.6 m3/min and 1.7 m3/min) in accordance with
the EM&A manual. The flow rate shall be indicated on the flow rate chart.
n The power
supply shall be checked to ensure the samplers worked properly.
n On sampling,
the samplers shall be operated for 5 minutes to establish thermal equilibrium
before placing any filter media at the designated air quality monitoring
station.
n The filter
holding frame is then removed by loosening the four nuts and carefully a
weighted and conditioned filter is centered with the stamped number upwards, on
a supporting screen.
n The filter
shall be aligned on the screen so that the gasket formed an airtight seal on
the outer edges of the filter. Then the filter holding frame is tightened to
the filter holder with swing bolts. The applied pressure should be sufficient
to avoid air leakage at the edges.
n The shelter
lid shall be closed and secured with the aluminum strip.
n The timer is
then programmed. Information shall be recorded on the record sheet, which
included the starting time, the weather condition and the filter number (the initial
weight of the filter paper can be found out by using the filter number).
n After
sampling, the filter shall be removed and sent to laboratory for weighing. The
elapsed time is also recorded.
n Before
weighing, all filters are equilibrated in a conditioning environment for 24
hours. The conditioning environment temperature should be between 25°C and 30°C
and not vary by more than ±3°C; the relative humidity (RH) should be < 50%
and not vary by more than ±5%. A convenient working RH is 40%. Weighing results
are returned to MCL for further analysis of TSP concentrations collected by
each filter.
2.2.2 1-hour TSP air quality monitoring
Operating / Analytical
Procedures
The
measuring procedures of the 1-hr dust meter are in accordance with the
Manufacturer’s instruction Manual as follows:
§ Pull up the air sampling inlet cover
§ Change the Mode 0 to BG once
§ Push Start/Stop switch once
§ Turn the knob to SENSI.ADJ and press it
§ Push Start/Stop switch once
§ Return the knob to the position MEASURE slowly
§ Push the timer set switch to set measuring time
§ Remove the cap and make a measurement
2.3.1 According to
the EM&A Manual, three air quality monitoring locations, namely KTD1, KTD2
and KER1, are covered by this Contract within the South Apron Area of Former Kai
Tak Airport. The other two air quality monitoring locations, which are
identified in Cha Kwo Ling area, are farther than 500m away from the site
boundary and thus not covered by this Contract. The monitoring works in Cha Kwo
Ling area are covered by Contract No. ED/2018/04.
2.3.2 According to
the approved alternative baseline air quality and noise monitoring locations
(EPD reference: EP2/K19/A/21 pt.5), the original monitoring locations (KTD1,
KTD2 and KER1) are proposed to be replaced by alternative monitoring locations
(KTD1a, KTD2a and KER1a) for air quality monitoring.
2.3.3 According to
the approved relocation of monitoring location KER1a (EPD reference: () in
EP2/K19/A/21 pt.5), the monitoring location KER1a are proposed to be relocated
by alternative monitoring location KER1b for air quality monitoring.
2.3.4 According to
the approved relocation of monitoring location KTD2a (EPD reference: () in
EP2/K19/A/21 pt.6), the monitoring location KTD2a are proposed to be relocated
by alternative monitoring location KTD2b for air quality monitoring.
2.3.5 According to
the approved relocation of monitoring location KTD2b (EPD reference: () in
EP2/K19/A/21 pt.7), the monitoring location KTD2b are proposed to be relocated
by alternative monitoring location KTD2c for air quality monitoring.
2.3.6 As informed by the ET of Contract No. ED/2018/04, the monitoring location
KTD1a and KER1b have been relocated to KTD1 and KER1 for air monitoring on 3
August 2020.
2.3.7 The
locations are summarized in Table 2.1 and shown in Figure 2.
Table 2.1 Location of Air Quality Monitoring
Station
Monitoring Station |
Location |
KTD1 |
Centre of Excellence in Paediatrics
(Rooftop of Children’s Hospital) |
KTD1a |
Centre of Excellence in Paediatrics
(Children’s Hospital) |
KTD2a |
G/IC Zone next to Kwun Tong
Bypass(Future Hospital at Site 3C1) |
KTD2b |
G/IC Zone next to Kwun Tong Bypass (Next to the site of the New Acute
Hospital) |
KTD2c |
G/IC Zone next to Kwun Tong Bypass
(Next to the Kowloon Bay Sewage Interception Station) |
KER1 |
Future Residential Development at
Kerry Godown |
KER1a |
Site Boundary at Cheung Yip Street |
KER1b |
Site Boundary at Cheung Yip Street, close to the open space car park
area |
Baseline Monitoring
2.4.1 The baseline
air quality monitoring results established was used for the Project (from 20
January 2016 to 3 February 2016) and derive the Action and Limit Levels. The
graphical presentations for baseline air quality monitoring over the project
period are shown in Appendix C.
2.4.2 The
graphical presentations for impact air quality monitoring in the reporting
period are shown in Appendix D.
2.4.3 Total 17
Action Level and 2 Limit Level exceedance were
recorded in the reporting period. Base on the investigation, the occasional exceedances were non-project
related. In general, the monitoring parameters were within the range of the
baseline levels.